High Pressure Reducer Units
You can examine the features of Central Gas Power Plants (Cylinder Feeding System) in our medical gas systems group below.
- The central gas plant consists of pressure tubes and two-stage high-pressure reducers, each connected by means of a collector
- Collectors are used for the transmission of pressurized gas in the tubes to the regulators. The tubes are fixed to the collectors with chain
- There is a valve at the end of the collector for high pressure relief
- Gas-specific flexible connection pipes are used between the manifold and the cylinders
- Product code (automatic) : PP-OPR-xyz
- Product Code (Manual) : PP-MPR-xyz
x: 0 (Oxygen), A (Nitrous Oxide), SA (Pure Nitrogen), AR (Argon), KG (Mixed Gas), K (Corbondioxide), HD (Hydrogen), H (Dry Air), HE (Helium), AT (Acetylene), EO (Ethylene Oxide)
y: Number of collectors (2, 4, 6, 8), z: Number of cylinders (1, 2, 3, 4, 5)
AUTOMATİC MANIFOLD SYSTEM
CYLINDER | OXYGEN | NITROUSOXIDE | |||
2 x 1 | PP-OPR-O21 | PP-OPR-A21 | |||
2 x 2 | PP-OPR-O22 | PP-OPR-A22 | |||
2 x 3 | PP-OPR-O23 | PP-OPR-A23 | |||
2 x 4 | PP-OPR-O24 | PP-OPR-A24 | |||
2 x 5 | PP-OPR-O25 | PP-OPR-A25 | |||
4 x 5 | PP-OPR-O45 | PP-OPR-A45 | |||
6 x 5 | PP-OPR-O65 | PP-OPR-A65 | |||
8 x 5 | PP-OPR-O85 | PP-OPR-A85 |
MANUAL MANIFOLD SYSTEM
CYLINDER | OXYGEN | NITROUSOXIDE | |||
1 x 1 | PP-MPR-O11 | PP-MPR-A11 | |||
1 x 2 | PP-MPR-O12 | PP-MPR-A12 | |||
1 x 3 | PP-MPR-O13 | PP-MPR-A13 | |||
1 x 4 | PP-MPR-O14 | PP-MPR-A14 | |||
1 x 5 | PP-MPR-O15 | PP-MPR-A15 |
Full Automatic High Pressure Reducer
- High Pressure Reducer Units reduces the pressure of gas come from collector which is
mounted to the cylinders.
The pressure of gas in collectors (150 bar —MPa —2176 psi) is reduced to 4-7 bar that is appropriate pressure to provide for patients and medical devices through pipeline systems - Supplies any type of medical gas from both left and right hand manifold racks.
The left or right manifold bank can be defined as "Mission" and the MC S will automatically supply the distribution system from the
"Standby " bank when pressure in the "Mission" bank drops to a pre-determined level. - Sytem includes a digital panel with visual and sound alarm.
(MES C11, HTM2022) - Product Code: PP-OPR-x
x: 0 (Oxygen), A (Nitrous Oxide), SA (Pure Nitrogen), AR (Argon), KG (Mixed Gas), K (Corbondioxide), HD (Hydrogen), H (Dry Air), HE (Helium), AT (Acetylene), PP-OPR-x
EO (Ethylene Oxide)
Manuel Yüksek Basınç Düşürücü
- High Pressure Reducer used as a backup system or used in manuel plants
- Product Code: PP-MPR-x
x: 0 (Oxygen), A (Nitrous Oxide), SA (Pure Nitrogen), AR (Argon), KG (Mixed Gas),
K (Corbondioxide), HD (Hydrogen), H (Dry Air), HE (Helium), AT (Acetylene),
PP-OPR-x
EO (Ethylene Oxide)