Medical Air Plants

It can be understood from the introduction of the Medical Air Plant, which is in our medical gas systems group.

  • It is a central system used to supply the medical air required for hospital departments. It is capable of providing a flow rate of 10 bar with one compressor not running (after dryer losses) (MES C11, HTM2022)
  • Depending on the project or system capacity, 2 or 3 compressors are used. Each can provide 10 bar air pressure (class F, BS 1999)
  • Vertical or horizontal air vessel is used depending on the project or system capacity. When using a horizontal tank, compressors can be mounted on the tank (EN 286, BS 5169/1975 Class 3 Grade E)
  • Air delivery is via duplex sets of filters, pre-filter, dust filter, activated carbon filter and bacterial filter (BS 7226:1989)
  • Duplex desiccant is used for air drying. It is controlled by a separate control panel and incorporate a dew point alarm system
  • Equal-time operation of each compressor is ensured by the automatic control panel (IEE, BS 4941)
  • 4-7 bar regulation group with by-pass allows air pressure to be adjusted according to the places to be used
  • MDD 93/42/EEC, Class IIb Medical Devices
PRODUCT CODE COMP. CAPACITY
PP-MAS-235 2 x 35 m³/h
PP-MAS-240 2 x 40 m³/h
PP-MAS-260 2 x 60 m³/h
PP-MAS-290 2 x 90 m³/h
PP-MAS-2110 2 x 110 m³/h
PP-MAS-2150 2 x 150 m³/h
PP-MAS-2200 2 x 200 m³/h
PP-MAS-2250 2 x 250 m³/h
PP-MAS-2300 2 x 300 m³/h
PRODUCT CODE COMP. CAPACITY
PP-MAS-335 3 x 35 m³/h
PP-MAS-340 3 x 40 m³/h
PP-MAS-360 3 x 60 m³/h
PP-MAS-390 3 x 90 m³/h
PP-MAS-3110 3 x 110 m³/h
PP-MAS-3150 3 x 150 m³/h
PP-MAS-3200 3 x 200 m³/h
PP-MAS-3250 3 x 250 m³/h
PP-MAS-3300 3 x 300 m³/h